Spectrum 44: BR1 Edit
DeleteMeasurement
Brand |
Namiba Terra Namiba Terra GmbH http://www.namibaterra.de/ |
---|---|
Lamp Product |
UV-Replux 15W |
Lamp ID |
BR1 (11/2007) |
Spectrometer | USB 2000 |
Ballast | - no ballast or default/unknown ballast - |
Reflector | |
Distance | 10 cm |
Age | 1 hours |
Originator (measurement) | Frances Baines |
Colorimetry
Colorimetry is the science to describe physically the human color perception. The wavelength range 380 nm - 780 nm is visible to humans and detected by three different photoreceptors. Many Reptiles see the range 350 nm - 800 nm and have an additional UV photoreceptor in their retina.
WARNING: The measurement range (350 - 800 nm) is not sufficient for this evaluation! Data is only available in the range 250.17 - 780.01 nm. Results are shown anyway but should be ignored by anyone except experts.
Whereas a spectrometer measures the intensity in every tiny wavelength interval resulting in thousands of individual intensities, the human eye only measures three intensities detected by the three cones. The same is true for the reptile eye with usually three or four photoreceptors. Effectively the detailled spectrum displayed above reduces to a much compacter bar graph displayed below. The photoreceptor sensitivites from these L-Cone, M-Cone, S-Cone, and U-Cone are used, they are chosen as an average of measured reptile photoreceptor sensitivity curves. The bar graph also shows as reference the intensity seen by the three or four photoreceptors for average sunlight (id 1).
From these three numbers the colour coordinate and the correlated colour temperature for humans are calculated using the CIE standard method. I adapted this concept to a "3 cone reptile (M,S,U)" and a "4 cone reptile (L,M,S,U)". I am sure, that this adaption to other colour spaces makes sense mathematically and this is also done in scientific research regarding colour vision of animals, however I have not seen calculation of colour temperatures for other animals in the scientific literature. Even if it is hypothetical, at least this shows, how arbitrary the colour temperature is, and that the colour temperature calculated for humans does not apply to reptiles. The colour spaces also show the colour coordinates of different phases of daylight ((ids 1, 338 – 451, 511 – 513 ), indicated by crosses, coloured in the appriximate colour perceived by a human.
Human (CIE) | 3 cone reptile | 4 cone reptile | |
---|---|---|---|
Cone Excitation | |||
Colour Coordinate | ( 0.29 ; 0.31 ) | ( 0.43 ; 0.43 ) | ( 0.26 ; 0.32 ; 0.32 ) |
CCT | 7800 Kelvin | 4600 Kelvin | 5400 Kelvin |
distance | 0.094 | 0.087 | |
colour space | 3-D-graph not implemented yet |
Vitamin D3 Analysis
Vitamin D3 is produced by UVB radiation around 300 nm. 7DHC/ProD3 present in the skin is converted to PreD3 when absorbing an UV photon. PreD3 can be converted back to ProD3, to Lumisterol, or to Tachysterol when absorbing another UV photon or can be converted to Vitamin D3 in a warm environment.
This process prevents any overdose of vitamin D3 from UV radiation with a spectrum similar to sunlight. As a comparison the solar spectra at 20°(id:14) and at 85°(id:21) solar angle are shown.
The ratio of the two solarmeters 6.2 (UVB) and 6.5 (UV index) readings has proven a useful and very simply number to acess the spectral shape in the vitamin-d3-active region.
Effective Irradiances
Effective irradiances are calculated for all ranges, actionspectra and radiometers currently present in this database.
The calculation method is a numerical implementation (Simpson's rule) of the formula
To learn more about calculating effective irradiances and radiometers I recommend this excellent report on UVB meters: Characterizing the Performance of Integral Measuring UV-Meters (pdf).
The numbers in the following tables can also be used to estimate certain (effective) irradiances from radiomer readings. Example: If the database lists
- range: UVB (US) = 13.8 µW/cm²
- radiometer: Solarmeter 6.2 = 19.6 µW/cm²
total ( 0 nm - 0 nm) 818 µW/cm² = 8.18 W/m² UVC ( 0 nm - 280 nm) 2 µW/cm² = 0.02 W/m² non-terrestrial ( 0 nm - 290 nm) 2.57 µW/cm² = 0.0257 W/m² total2 ( 250 nm - 880 nm) 818 µW/cm² = 8.18 W/m² UVB (EU) ( 280 nm - 315 nm) 25.7 µW/cm² = 0.257 W/m² UVB (US) ( 280 nm - 320 nm) 45.4 µW/cm² = 0.454 W/m² UVA+B ( 280 nm - 380 nm) 215 µW/cm² = 2.15 W/m² Solar UVB ( 290 nm - 315 nm) 25.1 µW/cm² = 0.251 W/m² UVA D3 regulating ( 315 nm - 335 nm) 86.6 µW/cm² = 0.866 W/m² UVA (EU) ( 315 nm - 380 nm) 189 µW/cm² = 1.89 W/m² UVA2 (medical definition) ( 320 nm - 340 nm) 92.5 µW/cm² = 0.925 W/m² UVA (US) ( 320 nm - 380 nm) 169 µW/cm² = 1.69 W/m² UVA1 (variant) ( 335 nm - 380 nm) 102 µW/cm² = 1.02 W/m² UVA1 (medical) ( 340 nm - 400 nm) 81 µW/cm² = 0.81 W/m² vis. UVA ( 350 nm - 380 nm) 36.6 µW/cm² = 0.366 W/m² VIS Rep3 ( 350 nm - 600 nm) 465 µW/cm² = 4.65 W/m² VIS Rep4 ( 350 nm - 700 nm) 609 µW/cm² = 6.09 W/m² purple ( 380 nm - 420 nm) 29.2 µW/cm² = 0.292 W/m² VIS ( 380 nm - 780 nm) 601 µW/cm² = 6.01 W/m² VIS2 ( 400 nm - 680 nm) 551 µW/cm² = 5.51 W/m² PAR ( 400 nm - 700 nm) 568 µW/cm² = 5.68 W/m² tmp ( 400 nm - 1100 nm) 597 µW/cm² = 5.97 W/m² blue ( 420 nm - 490 nm) 174 µW/cm² = 1.74 W/m² green ( 490 nm - 575 nm) 181 µW/cm² = 1.81 W/m² yellow ( 575 nm - 585 nm) 22.1 µW/cm² = 0.221 W/m² orange ( 585 nm - 650 nm) 102 µW/cm² = 1.02 W/m² red ( 650 nm - 780 nm) 92.6 µW/cm² = 0.926 W/m² IRA ( 700 nm - 1400 nm) 28.6 µW/cm² = 0.286 W/m² IR2 ( 720 nm - 1100 nm) 17 µW/cm² = 0.17 W/m² IRB ( 1400 nm - 3000 nm) 0 µW/cm² = 0 W/m²
Erythema 2.47 UV-Index Pyrimidine dimerization of DNA 16.1 µW/cm² Photoceratitis 3.28 µW/cm² Photoconjunctivitis 1.83 µW/cm² DNA Damage 2.22 Vitamin D3 6.84 µW/cm² Photosynthesis 403 µW/cm² Luminosity 1520 lx Human L-Cone 223 µW/cm² Human M-Cone 197 µW/cm² Human S-Cone 139 µW/cm² CIE X 198 µW/cm² CIE Y 211 µW/cm² CIE Z 263 µW/cm² PAR 2670000 mol photons Extinction preD3 66 e-3*m²/mol Extinction Tachysterol 210 e-3*m²/mol Exctincition PreD3 41300 m²/mol Extinction Lumisterol 22.8 m²/mol Exctincition Tachysterol 302000 m²/mol Extinction 7DHC 25.5 m²/mol L-Cone 184 µW/cm² M-Cone 229 µW/cm² S-Cone 229 µW/cm² U-Cone 72.8 µW/cm² UVR - ICNIRP 2004 3.04 Rel Biol Eff Melatonin Supression 202 µW/cm² Blue Light Hazard 155 µW/cm² (102 µW/cm² per 1000 lx) CIE 174:2006 PreVit D3 7.37 µW/cm² Lumen Reptil 1650 "pseudo-lx" Vitamin D3 Degradation 9.45 µW/cm² Actinic UV 3.02 µW/cm² (19.9 mW/klm) Exctincition Lumisterol 26900 m²/mol Exctincition 7DHC 27800 m²/mol Exctincition Toxisterols 12600 m²/mol
Solarmeter 6.2 (UVB, pre 2010) 58.1 µW/cm² Solarmeter 6.5 (UV-Index, pre 2010) 2.16 Leybold UVB 38.8 µW/cm² Leybold UVA 121 µW/cm² Leybold UVC 1.48 µW/cm² DeltaOhm UVB 105 µW/cm² DeltaOhm UVC 13.8 µW/cm² Vernier UVB 11.9 µW/cm² Vernier UVA 143 µW/cm² Gröbel UVA 158 µW/cm² Gröbel UVB 19.2 µW/cm² Gröbel UVC 1.57 µW/cm² Luxmeter 1560 lx Solarmeter 6.4 (D3) 6.75 IU/min UVX-31 113 µW/cm² IL UVB 0.0246 µW/cm² IL UVA 132 µW/cm² Solarmeter 6.5 (UVI, post 2010) 1.79 UV-Index Solarmeter 6.2 (UVB, post 2010) 39.3 µW/cm² (Solarmeter Ratio = 22) Solarmeter AlGaN 6.5 UVI sensor 22.8 UV Index GenUV 7.1 UV-Index 1.38 UV-Index Solarmeter 10.0 (Global Power) (manuf.) 7.13 W/m² Solarmeter 4.0 (UVA) 1.68 mW/cm² LS122 (manuf.) 0 W/m² ISM400 (first guess) 4.51 W/m² LS122 (assumption) 0.209 W/m² ISM400_new 3.68 W/m² Solarmeter 10.0 (Global Power) (assumption) 6.39 W/m²